The Use of
Known Discontinuity Standards
by Israel Vasquez*
| How
do you know when your fluorescent penetrant testing materials
are no longer performing as they should? You test them, of course!
This month's author provides some insight into how this testing
of the materials should be done. It is basic to nondestructive
testing that we test our testing materials so that we can do the
job that we think we are doing.
Frank Iddings
Tutorial Projects Editor
|
Figures
1-3
The
intent of this paper is to provide some insight for understanding the
requirements for conducting the system performance test of a fluorescent
penetrant testing system. This author has encountered a great deal of
confusion throughout the aerospace nondestructive testing (NDT) industry
as to what is the proper method for conducting this test. The uncertainties
this paper will address are those concerning the system performance
test and the proper acceptance criteria. In addition, it is intended
to clarify the confusion between the system performance test and the
penetrant sensitivity test.
The system performance test required
by specifications such as MIL-STD-6866 (1996), ASTM E 1417 (1999)
and SAE AMS 2647 (1995) require that a known discontinuity standard
be processed through the penetrant testing system using standard parameters
and that the indications detected be compared to those obtained with
unused penetrant materials (Vasquez, 1997). This comparison is to be
made with photographs (not recommended by this author), other such records
of previously obtained indications or with a similar known discontinuity
standard processed with the unused materials. Known discontinuity standards
generally used by the aerospace industry for conducting this test are
the penetrant system monitor (PSM-5) panel, the TAM test panel and recently
the five star penetrant test panel. All three are recognized throughout
the industry as the test panel. For all intents and purposes, the term
"test panel" shall be used throughout this paper (see
Figure 1).
An
acceptable test would convey confidence that the system is in
control to the user of the penetrant testing system
The failure mode was disbonding of
the cardboard from the steel strip. The transmission lost torque. The
problem was posed to my NDT group as a requirement to develop a test
for adhesion of the cardboard to the steel band.
For several years, several aerospace
specifications specified imprecise acceptance criteria when conducting
the system performance test using the test panels. These specifications
required that, on a daily basis, the test panel exhibit a specified
number of indications based on the penetrant's sensitivity (see
Table 1).
In addition, several aerospace specifications
acknowledged this test as a penetrant sensitivity test. This only added
to the quandary that we are experiencing in the industry today. The
acceptance criteria prescribed by these specifications do not take into
account that the methods for inducing the cracks of these panels were
not collaborated with empirical data (for example, comparing different
penetrant sensitivities and manufacturers against a particular number
of known crack sizes). In addition, there are several manufacturing
inconsistencies, such as plating thickness and varying forces used by
the hardness indenter, that contribute to the test panel being an imprecise
and inappropriate tool for determining a penetrant's sensitivity. The
fundamental objective of the system performance test is for the penetrant
testing system to be able to demonstrate its capability to reproduce
the same number of indications of the known discontinuity standard on
a continual basis regardless of the penetrant's sensitivity level. An
acceptable test would convey confidence that the system is in control
to the user of the penetrant testing system. Note that the penetrant
testing system encompasses as a minimum the penetrant materials, washing
equipment, lighting equipment, temperatures, pressures and the operator's
techniques (an essential element). Therefore, when using the test panel,
the system performance test is essentially a test by attributes (go/no
go), not a test by variables.
In essence, the criteria for the penetrant
testing system in reality should be stated as follows: the number of
indications found during the system performance test shall be equal
to those found during the initial qualification of the penetrant testing
system with all parameters, equipment and techniques being equal.
If the system is not able to exhibit
the same number of indications found during the initial qualification,
then the panel should be thoroughly cleaned and reprocessed. If after
reprocessing, the required number of indications is still not evident,
then a referee (master) panel for the same level of penetrant sensitivity
and method of removal (for example, water washable or postemulsifiable)
should be processed and its indications compared to its respective initial
indications. Note that panels should not be compared against each other,
since no two panels are alike. If the master panel fails to display
the required number indications, then an investigation and subsequent
corrective action is warranted for the system.
Compliance
Even though most of the aerospace industry requires the use of a known
discontinuity standard for the daily system performance test, a number
of them, remarkably, do not require a penetrant system/test panel qualification
(an initial calibration). Consequently, when conducting an audit of
a laboratory's fluorescent penetrant testing system, NDT auditors need
to be aware of this disparity and should therefore conduct a through
review of the pertinent specification(s) prior to the audit.
Although most aerospace specifications,
including MIL-STD-6866, ASTM E 1417 and SAE AMS 2647,
require that a comparison of the indications obtained on a daily basis
to those originally obtained with unused samples of the same materials
be employed as the acceptance criteria, the industry's approach has
been to compare the number of indications to the penetrant's sensitivity
levels (see Table 1). It is this author's assertion
that the criterion has always been properly stated in MIL-STD-6866
and ASTM E 1417, but, ever since the initiation of the test
panel, it has been inappropriately mandated by the aerospace industry,
misunderstood by NDT laboratories/facilities and mistakenly accepted
by NDT auditors.
| Table
1 Test panel specified minimum number of indications based
on penetrant's sensitivity |
|
| Sensitivity Level |
Minimum Number of Indications |
| Low |
Two largest |
| Normal |
Three largest |
| High |
Four largest |
| Ultra high |
All five |
|
Penetrant Sensitivity Test
Certain aerospace primary specifications, MIL-STD-6866
and ASTM E 1417 require NDT facilities to compare unused versus
in use penetrant materials on either a monthly or weekly basis by
processing both materials in order to demonstrate that there is no
significant difference between them. The acceptance criteria referenced
by these specifications is "sensitivity of the in-use penetrant noticeably
less than the reference (un-used) is unsatisfactory." This author's
opinion is that noticeable difference should mean not only the number
of indications seen, but also the size and brilliance of the indications.
Although the specifications require
the sensitivity test to be conducted using the same procedures used
for conducting the system performance test, they do not reference
the type of known discontinuity standard to be utilized. Many NDT
facilities, including certain penetrant testing laboratories, are
using test panels for conducting the sensitivity test.
It is the opinion of this author
that it is not reasonable to use test panels for the penetrant sensitivity
test unless the criterion specified is to simply count the number
of indications found and compare this number against those found with
unused materials or (recommended) each indication is measured and
compared to original measurements (within an appreciable tolerance).
Surprisingly enough, test panels may be used for conducting the penetrant
sensitivity test, if they are properly validated.
Rationale
During the initial qualification (upon receipt) of the test panel
used for the system performance test, the laboratory should process
it with unused samples of the same materials to be used with the penetrant
testing system. This author recommends that the test panels be processed
a minimum of three times for continuity during this initial qualification.
Each indication should then be measured and the average size recorded.
It stands to reason that if a laboratory periodically reprocesses
the panel with in use penetrant (unused emulsifier, when applicable)
and unused developer, then measures the indications and compares this
data against the original sizes (within an appreciable tolerance),
the lab is meeting the intent of the specifications for the penetrant
sensitivity test. Of course, this is only possible with a properly
maintained known discontinuity standard.
In addition to the sensitivity test,
these measurements may also be used for conducting the periodic maintenance
(degradation) checks of the test panels required by certain aerospace
corporations' specifications and industry specifications. ASTM
E 1417-99, paragraph 7.8.3.1 states "The maintenance procedures
shall ensure that cleaning of the standards between usages is adequate
and that physical changes in the standard that make it unsuitable
for use can be detected." By periodically reprocessing and remeasuring
the indications, one is ensuring that physical changes in the standard
that make it unsuitable for use is being detected.
Ideally, the preferred method for
conducting the penetrant sensitivity test would be to use a known
discontinuity standard which provides a more accurate test (side by
side comparison under the black light) such as the NiCr twin panels
(see Figure 2) or the
twin known discontinuity standard panels (see
Figure 3).
One of the panels is to be processed
with in use penetrant material (unused emulsifier, when applicable)
and unused developer. The accompanying panel is processed with unused
penetrant material, (unused emulsifier, when applicable) and unused
developer. Both panels are then observed side by side under the proper
ultraviolet light for overall brightness, color, presence and/or absence
of the indications. Any noticeable difference between the materials
shall be cause for rejection of the in use penetrant material. Note
that photographs of the known discontinuity standard's indications
are not reliable, primarily due to the fact that the photo must be
viewed under a white light while the known discontinuity standards
are viewed under an ultraviolet light and, therefore, it is not an
actual side by side comparison.
Conclusion
Aerospace companies, laboratories and auditors should be advised to:
 |
verify that the
users qualify (calibrate) all test panels with their respective
penetrant testing system immediately after purchasing to confirm
the quantity and actual sizes of the starburst/linear indications |
|
demonstrate repeatability
on a daily basis of the same number of indications found during
the initial penetrant testing system/test panel qualification
(calibration) regardless of the penetrant's sensitivity level |
|
for the periodic
sensitivity test, use test panels only when each indication size
is known and this data used for comparison against in use materials
(original measurements) |
|
or as an alternate,
require the use of NiCr twin panels or the twin known discontinuity
standard panels for the sensitivity test. |
Acknowledgments
Pertinent information used herein was provided by Amos Sherwin
and Sam J. Robinson of Sherwin, Inc.
References
Army Materials Directorate, MIL-STD-6866, Military Standard
Inspection Liquid Penetrant, Aeronautical Systems Center, 1996.
ASTM International, ASTM E 1417,
Standard Practice for Liquid Penetrant Examination, Annual Book
of ASTM Standards, Metals Test Method and Analytical Procedures,
Vol. 03.03, 2000, pp. 726-735.
Society of Automotive Engineers,
SAE AMS 2647A, Fluorescent Penetrant Inspection Aircraft and Engine
Component Maintenance, 1995.
Vasquez, Israel, "Fluorescent Penetrant
Inspection System Performance Test Using Known Defect Standards,"
Materials Evaluation, Vol. 56, 1997, pp. 1319-1322.
* Vastek Consulting,
7290 Jasmine Dr., Hanover Park, IL 60133; (630) 213-3432; fax (630)
213-3495; e-mail <vastek@aol.com>.
Copyright ©
2002 by the American Society for Nondestructive Testing, Inc. All
rights reserved.
[ Materials
Evaluation ]
Copyright © 2010 by the American Society for Nondestructive Testing, Inc. ASNT is not responsible for the authenticity or accuracy of information herein. Published opinions and statements do not necessarily reflect the opinion of ASNT. Products or services that are advertised or mentioned do not carry the endorsement or recommendation of ASNT.
IRRSP, NDT Handbook, The NDT Technician and www.asnt.org are trademarks of the American Society for Nondestructive Testing, Inc. ACCP, ASNT, Level III Study Guide, Materials Evaluation, Nondestructive Testing Handbook, Research in Nondestructive Evaluation and RNDE are registered trademarks of the American Society for Nondestructive Testing, Inc. ASNT exists to create a safer world by promoting the profession and technologies of nondestructive testing.
|